A negative dimension defining source concentration is present in the hot spot template parameters. The manual's first action is: “Make sure that all parameters defining the source concentration are positive, then try again.”
- Model
- Canberra 9233653E V3.0
- Display shows
- #602
- Type
- Error code
- Source
- Manual p. 424
What #602 means
AN INVALID DIMENSION IN THE RECTANGULAR HOT SPOT. CHECK THE HOT SPOT DIMENSIONS IN SOURCE TABLE. p. 424
Common causes
- Upon checking the 'complex box' template dimensions, it was found that some dimension defining the source concentration is negative. p. 424
What to do
- Make sure that all parameters defining the source concentration are positive, then try again. p. 424
Before you service the product, read the safety instructions in the manual p. 526.
If #602 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Canberra service.
This page summarises what the manufacturer’s manual says about #602; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Canberra. Found a mistake? Report an error.