User manual for Oakton's pH, ORP, and TDS continuous monitors covering calibration, operation, error messages, and accessories.
This manual provides operating instructions for Oakton’s pH, ORP, and TDS continuous monitors. It covers calibration procedures for each type, including pH (3-point), ORP (offset adjustment), and TDS (conductivity). The monitors feature microprocessor technology for user-friendly operation and are designed for long-term, low-maintenance use. Calibration is required for accuracy and should be performed regularly. The manual also includes error messages, specifications, and a list of accessories such as replacement electrodes and calibration solutions.
| pH Range | 0.0 to 14.0 pH |
|---|---|
| pH Resolution | 0.1 pH |
| pH Accuracy | ±0.1 pH |
| Display | LED, 3 digits |
| Power | 110 VAC |
| Operating Temperature | 0 to 50°C |
| Storage Temperature | –10 to 60°C |
| Humidity | 10 to 95% RH (non condensing) |
| Dimensions (Monitor) | 3.1" L x 3.4" W x 1.25" H |
Plug in the monitor, rinse the probe, dip it in the calibration buffer, press CAL, wait for the reading to stabilize, then press ENTER.
The ORP monitor has an offset adjustment of ±100 mV in 5 mV increments.
The manual lists replacement electrodes, calibration solutions, and rinse water solution pouches.
The product is warranted for three years from the date of purchase, excluding the electrode.