This is a portable pH tester that can measure pH, ORP, EC and TDS parameters. It can be used for water quality testing, agriculture, food and other industries.
This manual explains how to operate the Oakton pHTestr® 30 pocket pH tester, including turning it on/off, calibrating with 1-3 points, and measuring pH values. It also covers probe preparation, storage, and troubleshooting. The device features auto buffer recognition, IP67 waterproof rating, and a replaceable double-junction sensor. Calibration requires specific buffer solutions, and the tester will not function accurately with distilled water.
| Accuracy | ±0.01 pH ± 1 digit |
|---|---|
| Calibration Points | 1, 2, or 3 points |
| Auto Buffer Recognition | US or NIST |
| Waterproof Rating | IP67 |
| Sensor Type | Replaceable double-junction sensor |
Short press the /MEAS key to turn on; long press to turn off.
No, storing the probe in distilled or deionized water causes permanent damage. Use 3M KCL storage solution instead.
The tester has an IP67 waterproof rating.