Oakton pHTestr® 30 Pocket pH Tester Operating Manual

This is a portable pH tester that can measure pH, ORP, EC and TDS parameters. It can be used for water quality testing, agriculture, food and other industries.


Oakton pHTestr® 30 Pocket pH Tester Operating Manual - cover page
Brand
Oakton
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
319 KB
Published
19 October, 2023
Updated
19 October, 2023
MD5 checksum
6F5D8AF43D2A397EEBE4E5E8FEB09AF6

About this manual

This manual explains how to operate the Oakton pHTestr® 30 pocket pH tester, including turning it on/off, calibrating with 1-3 points, and measuring pH values. It also covers probe preparation, storage, and troubleshooting. The device features auto buffer recognition, IP67 waterproof rating, and a replaceable double-junction sensor. Calibration requires specific buffer solutions, and the tester will not function accurately with distilled water.

What's inside

  • Keypad Functions
  • Preparation Before Use
  • Calibration
  • pH Measurement
  • Setting the Parameters
  • Self-Diagnostic Information
  • Notes and Special Applications

Specifications

Accuracy±0.01 pH ± 1 digit
Calibration Points1, 2, or 3 points
Auto Buffer RecognitionUS or NIST
Waterproof RatingIP67
Sensor TypeReplaceable double-junction sensor

Frequently asked questions

How do I turn on and off the pHTestr® 30?

Short press the /MEAS key to turn on; long press to turn off.

Can I store the probe in distilled water?

No, storing the probe in distilled or deionized water causes permanent damage. Use 3M KCL storage solution instead.

What is the IP rating of the tester?

The tester has an IP67 waterproof rating.

Download PDF (319 KB)

More Oakton Manuals


View all Oakton manuals →