MRC WBH-100I/200I Circulation Water Baths Operation Manual

Operation manual for MRC WBH-100I/200I circulation water baths, covering installation, controls, operation, and maintenance.


MRC WBH-100I/200I Circulation Water Baths Operation Manual - cover page
Brand
MRC
Category
Other device
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
307 KB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
9227E0841DE856748B81E9AE5383AF04

About this manual

The MRC WBH-100I/200I circulation water baths are for professional, industrial, or educational use where materials are prepared or tested at approximately atmospheric pressure. This operation manual covers safety, receiving, installation, controls, and operation of the units.

It describes distilled-water-only use, temperature control and calibration, bath-cover requirements, and environmental and power conditions. The manual also includes handling, cleaning, and decontamination guidance.

What's inside

  • Introduction
  • General Safety Considerations
  • Receiving Your Unit
  • Installation
  • Controls and Components Overview
  • Operation
  • Turning On the Unit
  • Calibrating the Main Temperature Control

Specifications

Intended useProfessional, industrial, or educational use at approximately atmospheric pressure
Room temperature range18 to 40 C
Relative humidityNo greater than 80% at 25 C
Supply voltage toleranceWithin ±10% of data plate rating
Power frequency50/60 Hz
Minimum clearance10 cm from walls or partitions
Water typeDistilled water only
Normal water depth5 1/2 inches (14 cm)
Minimum water depth4.7 inches (12 cm) over rack bottom
Bath coverRequired to reach set points above 60 C

Frequently asked questions

What water should be used in the bath?

Use distilled water only; do not use tap water, deionized water, or chemicals.

Can the bath be used as an acid bath?

No. These baths are not intended for use as acid baths, and such use can cause severe damage and void the warranty.

When is the bath cover required?

The supplied bath cover must be used to reach set points above 60 C.

What is the normal water depth?

Normal depth is 5 1/2 inches (14 cm), and it must be at least 4.7 inches (12 cm) over the bottom of the rack.

Download PDF (307 KB)