MRC STE-V-60/80 Vertical Autoclave Operation Manual

This manual provides instructions for the use of autoclave


MRC STE-V-60/80 Vertical Autoclave Operation Manual - cover page
Brand
MRC
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
37
File format
PDF
File size
1 MB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
4E46F0E4CC87BA0489AB839A9A9ED07D

About this manual

The MRC STE-V-60 and STE-V-80 are vertical autoclaves that use high-temperature, pressure-saturated steam generated by electrical heating. By controlling steam temperature and time, they denature protein and kill microorganisms for sterilization of items such as surgical instruments and glassware in medical units and laboratories.

This manual covers unloading, installation and movement, site, power and water requirements, storage and working conditions, technical parameters, parts, operation panel settings, operation and running, maintenance, inspection and repair, and wiring and PID drawings.

What's inside

  • General introduction
  • Basic introduction
  • Technical parameters
  • Parts introduction
  • Operation panel
  • Operation and running
  • Maintenance
  • Inspection and repairing
  • Appendix 1 wiring connection drawing
  • Appendix 2 PID drawing

Specifications

ModelSTE-V-60, STE-V-80
Power supplySingle phase 220V±10%, 50 Hz; three phase 220V±10%, 50 Hz
Total power5 kW (STE-V-60), 5.3 kW (STE-V-80)
Overall dimensionsW476×L616×H990 (STE-V-60); W546×L688×H1030 (STE-V-80)
Weight65 kg (STE-V-60), 75 kg (STE-V-80)
Volume50 L (STE-V-60), 80 L (STE-V-80)
Chamber sizeφ316×D667 (STE-V-60); φ386×D695 (STE-V-80)
Chamber material06Cr19Ni10 (stainless steel)
Sterilization temperature105°C to 138°C
Melt temperature60°C to 100°C (special designed)
Holding temperature40°C to 60°C
Temperature displayDigital display (0 to 150°C)
Pressure gaugeAnalog display (0 to 0.4 MPa)
Design pressure0.28 MPa
Sterilization time0 to 99 hours
Temperature holding time0 to 99 hours
AccessoriesStainless steel sterilization basket

Frequently asked questions

What water should be used in the autoclave?

Use soft water or pure water.

What are the storage environment requirements?

Store the autoclave at -20°C to 55°C with relative humidity less than 80%.

What single-phase power supply is suggested?

Single-phase AC 220V ±10% (50 Hz), 25 A or more.

What sterilization temperature range is supported?

The sterilization temperature range is 105°C to 138°C.