This manual provides instructions for the use of autoclave
The MRC STE-V-60 and STE-V-80 are vertical autoclaves that use high-temperature, pressure-saturated steam generated by electrical heating. By controlling steam temperature and time, they denature protein and kill microorganisms for sterilization of items such as surgical instruments and glassware in medical units and laboratories.
This manual covers unloading, installation and movement, site, power and water requirements, storage and working conditions, technical parameters, parts, operation panel settings, operation and running, maintenance, inspection and repair, and wiring and PID drawings.
| Model | STE-V-60, STE-V-80 |
|---|---|
| Power supply | Single phase 220V±10%, 50 Hz; three phase 220V±10%, 50 Hz |
| Total power | 5 kW (STE-V-60), 5.3 kW (STE-V-80) |
| Overall dimensions | W476×L616×H990 (STE-V-60); W546×L688×H1030 (STE-V-80) |
| Weight | 65 kg (STE-V-60), 75 kg (STE-V-80) |
| Volume | 50 L (STE-V-60), 80 L (STE-V-80) |
| Chamber size | φ316×D667 (STE-V-60); φ386×D695 (STE-V-80) |
| Chamber material | 06Cr19Ni10 (stainless steel) |
| Sterilization temperature | 105°C to 138°C |
| Melt temperature | 60°C to 100°C (special designed) |
| Holding temperature | 40°C to 60°C |
| Temperature display | Digital display (0 to 150°C) |
| Pressure gauge | Analog display (0 to 0.4 MPa) |
| Design pressure | 0.28 MPa |
| Sterilization time | 0 to 99 hours |
| Temperature holding time | 0 to 99 hours |
| Accessories | Stainless steel sterilization basket |
Use soft water or pure water.
Store the autoclave at -20°C to 55°C with relative humidity less than 80%.
Single-phase AC 220V ±10% (50 Hz), 25 A or more.
The sterilization temperature range is 105°C to 138°C.