This manual provides detailed instructions on how to use the steam sterilizer, including installation, operation and maintenance. Please read it carefully.
This is the operation manual for the MRC steam sterilizer in the 60 L and 80 L versions (STE-HT-60/80). It describes the product's structure — body, door, pipeline system and control system — and explains that it uses saturated damp and hot steam under high temperature and pressure to sterilize medical instruments, laboratory ware, culture medium and unsealed liquids.
The manual walks through unpacking and installation requirements, levelling, power and water supply, checking the accessories and commissioning the equipment. It then covers the keys and display screens, parameter setting, the sterilization programs and the step-by-step sterilization procedure (washing, packing, placement, running and taking out items). A maintenance section deals with cleaning, door gasket and filter replacement, tank drainage, printer paper, the safety valve and the fuse, and the closing chapters list common problems, alarm codes with treatment measures and the standard accessories.
| Model and volume | STE-HT-60: 60 L; STE-HT-80: 80 L |
|---|---|
| Usable chamber dimension | φ385×595 mm (60 L); φ385×717 mm (80 L) |
| Power supply | AC 220V/230V, 50/60 Hz |
| Power | 7.4 kVA |
| Design pressure | -0.1/0.30 MPa |
| Design temperature | 144 °C |
| Vacuum low limit | -0.08 MPa (Type B) |
| Temperature selection range | 105~138 °C |
| Temperature display precision | 0.1 °C |
| Pressure display precision | 1 kPa |
It is used for sterilization of medical instruments, laboratory ware, culture medium and unsealed liquid, as well as any material that may contact blood or body fluid.
The product uses damp and hot saturated steam at high temperature and pressure as its sterilization factor.
No. The manual states the product is not applicable for unsealed bottled liquid, and sterilizing a bottle with a tight seal may cause an explosion.
It means the chamber pressure exceeds the preset pressure; the manual says to check the pressure sensor.