This manual describes the operation of the ASB-CC22-VV2 series analytical balance.
The MRC ASB-C2-V2 Series Operation Manual describes the setup and use of the ASB-C2-V2 analytical balances. It covers technical data, transport and storage, unpacking and assembly, keyboard and user menu, weighing, balance adjustment, databases, working modes, communication, peripherals, and error messages.
The manual explains intended use in a laboratory environment, warns against dynamic weighing and magnetic loads, and provides procedures for zeroing, taring, internal and external adjustment, printouts, RS 232 and WIFI settings, and under hook weighing.
| ASB-60/220-C2-V2 maximum capacity | 60/220 g |
|---|---|
| ASB-C2-V2 series readability | 0.01/0.1 mg |
| ASB-C2-V2 series minimum capacity | 1 mg for ASB-60/220-C2-V2; 10 mg for other listed ASB-C2-V2 models |
| Working temperature | +10 °C to +40 °C |
| Power supply | 12 to 16 V DC / 2.1 A |
| ASB-C2-V2 pan size | 85 mm diameter |
It is used to determine the weighing value of loads in a laboratory environment as a non-automatic weighing instrument.
The listed power supply is 12 to 16 V DC / 2.1 A.
Send the command C1 CR LF. A response of C1_A CR LF followed by MASS FRAME means the command was understood and is in progress.
It means the value is beyond the zero range.