The laboratory furnace is a device for analyzing various materials and different types of thermal treatment.
The MRC MSF 12-4 is an electric laboratory furnace intended for analysis of various materials and different kinds of thermal treatment at temperatures up to 1100°C. It has a vacuumed fiber chamber with heating elements around the working chamber, a door with thermo insulation material, and an electronic temperature controller operating with a thermoelectric converter mounted in the chamber.
The manual describes special hazards, the complement, layouts and operation, exploitation instructions, mounting, preoperational procedure, operation, maintenance, storage, transportation, guarantee period, technical data, failures and removal methods, testing record, and the electric diagram. It states that the furnace has a 230 V power supply at 50 Hz and is designed for indoor use.
| Rated power | 1.8 kW |
|---|---|
| Rated supply voltage | 230 V |
| Rated frequency | 50 Hz |
| Number of phases | 1 |
| Rated temperature | 1100 °C |
| Heat-up time (without charge) | 50 min |
| Furnace temperature stability at rated temperature without charge | not more than ±2 °C |
| Furnace temperature uniformity at rated temperature without charge | not more than ±10 °C |
| Working chamber width | 200 mm |
| Working chamber depth | 300 mm |
| Working chamber height | 133 mm |
| Outside width | 440 mm |
| Outside depth | 620 mm |
| Outside height | 510 mm |
| Weight net | not more than 28 kg |
It is intended for analysis of various materials and different kinds of thermal treatment at temperatures up to 1100°C.
The furnace has a 230 V power supply at 50 Hz frequency.
The moisture must be expelled from the furnace before the first heating cycle with load or after a long unused period.
The guaranteed operating period is 12 months after sales, but not more than 24 months from the date of manufacture.