This is a manual for the use of an anaerobic chamber. The manual introduces the use and safety of the anaerobic chamber.
This document is the official Installation and Operation Manual for the MRC BACTRON series of anaerobic chambers, specifically models BACTRONEZ-2, BACTRON300-2, BACTRON600-2, and BACTRON900-2. It covers safety requirements, receiving and inspecting the unit, detailed installation steps including gas supply and leveling, and comprehensive operational procedures for creating and maintaining an anaerobic atmosphere.
The guide also details daily maintenance routines, troubleshooting common issues like persistent oxygen or leaks, and specific care for components such as the O2 scrubber cartridges and acrylic glass panels. It includes specifications for the units and a parts list for ordering consumables.
| Operating Pressure Range | 22.14 – 31.3 inHg (75 – 106 kPa) |
|---|---|
| Intended Use | Professional, industrial, or educational use |
| Gas Requirement | 5% hydrogen, 5% carbon dioxide, 90% nitrogen |
| O2 Scrubber Reactivation | 8 hours at 200°C |
The manufacturer recommends an AMG mixture ratio of 5% hydrogen, 5% carbon dioxide, and 90% nitrogen.
The BACTRON uses a minimum of one Oxoid brand oxygen-detecting anaerobic indicator strip every 24 hours.
Support hours are 6 am – 4:30 pm Pacific Coast Time, Monday through Friday.