Japanese guide for the IDEXX SediVue Dx urine sediment analyzer covering bacteriuria result interpretation, the bacterial confirmation kit, and dried smear…
This two-page Japanese guide from IDEXX Laboratories explains how to use the SediVue Dx urine sediment analyzer to examine urine for bacteria. It describes how bacterial results are reported (not detected, suspected to be present, or present), how to interpret semi-quantitative results and images alongside the patient's clinical signs and history, and what to do when a crystalline debris detected message appears.
The guide also gives step-by-step instructions for the SediVue bacterial confirmation kit used with the IDEXX VetLab Station, a procedure for preparing dried urine sediment smears for staining with Diff-Quik or a similar stain, and background notes on bacteriuria with cited references. This manual is written in Japanese.
Bacterial results are reported as "not detected", "suspected to be present", or "present".
165 µL of well-mixed urine is dispensed into a new tube, and later 165 µL of the mixed sample is injected into the cartridge.
After dispensing 165 µL of urine, add one drop of R1 reagent (red cap) and mix by inverting five times, then add one drop of R2 reagent (blue cap), mix again, inject 165 µL into the cartridge and press start.
After the slide is well dried, it is stained with a common smear stain such as Diff-Quik.