Hanna Instruments HI981030 Soil pH Tester Instruction Manual

Hanna Instruments HI981030 soil pH tester instruction manual covering calibration, electrode refilling and cleaning, battery replacement, auto-off and…


Hanna Instruments HI981030 Soil pH Tester Instruction Manual - cover page
Brand
Hanna Instruments
Category
sensor
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
725 KB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
598A1767FD9597F464271271DA021D08

About this manual

This instruction manual for the Hanna Instruments HI981030 Soil pH Tester explains how to operate, calibrate and maintain the meter so that measurements stay accurate. It describes one- and two-point calibration with pH 4.01 and pH 7.01 buffers, the display tags shown during calibration, and the steps of the direct ground measurement procedure, including preparing the soil hole with the HI721319 soil auger.

The manual also covers electrode care: rinsing, refilling the reference well with HI9071 gelled bridge electrolyte, soaking the probe in storage solution, and the cleaning solutions recommended for soil and humus deposits. Battery replacement (CR2032), the low battery and "Err" messages, the auto-off settings, the list of accessories, the one-year warranty statement and the full specifications are also included.

What's inside

  • Operation
  • Meter Calibration
  • One or Two-Point Calibration with pH 7.01
  • One-Point Calibration with pH 4.01
  • Procedure for Direct Ground Measurement
  • Probe Features
  • Refilling the Electrode
  • Care and Maintenance
  • Battery Replacement
  • Auto-off

Specifications

Range0.0 to 14.0 pH
Resolution0.1 pH
Accuracy±0.2 pH @25 °C/77 °F
CalibrationAutomatic, one or two-point
Battery typeCR2032 Li-ion
Battery lifeApproximately 1000 hours of continuous use
Auto-off8 minutes, 60 minutes or can be disabled
Environment0 to 50 °C (32 to 122 °F); RH 95% max
Dimensions51 x 151 x 21 mm (2 x 5.9 x 0.9 in.)
Weight44 g (1.6 oz.)

Frequently asked questions

How do I calibrate the Hanna HI981030 soil pH tester?

From measurement mode, press and hold the ON/OFF button until "CAL" is displayed, then place the probe tip in pH 4.01 or 7.01 buffer. When the reading is stable, "Sto" is displayed and the calibration is saved.

What battery does the meter use and how is it replaced?

It uses a CR2032 Li-ion battery. Turn the battery cover on the back of the meter counterclockwise to unlock, remove the cover and fit the new battery with the + side facing up.

What does the "Erb" message mean?

It appears when the battery has been depleted; the meter then turns off.

How often should the electrode be recalibrated?

The manual states that for best results the meter should be recalibrated periodically, and it should always be calibrated in fresh buffers before use.