Hanna Instruments HI981030 soil pH tester instruction manual covering calibration, electrode refilling and cleaning, battery replacement, auto-off and…
This instruction manual for the Hanna Instruments HI981030 Soil pH Tester explains how to operate, calibrate and maintain the meter so that measurements stay accurate. It describes one- and two-point calibration with pH 4.01 and pH 7.01 buffers, the display tags shown during calibration, and the steps of the direct ground measurement procedure, including preparing the soil hole with the HI721319 soil auger.
The manual also covers electrode care: rinsing, refilling the reference well with HI9071 gelled bridge electrolyte, soaking the probe in storage solution, and the cleaning solutions recommended for soil and humus deposits. Battery replacement (CR2032), the low battery and "Err" messages, the auto-off settings, the list of accessories, the one-year warranty statement and the full specifications are also included.
| Range | 0.0 to 14.0 pH |
|---|---|
| Resolution | 0.1 pH |
| Accuracy | ±0.2 pH @25 °C/77 °F |
| Calibration | Automatic, one or two-point |
| Battery type | CR2032 Li-ion |
| Battery life | Approximately 1000 hours of continuous use |
| Auto-off | 8 minutes, 60 minutes or can be disabled |
| Environment | 0 to 50 °C (32 to 122 °F); RH 95% max |
| Dimensions | 51 x 151 x 21 mm (2 x 5.9 x 0.9 in.) |
| Weight | 44 g (1.6 oz.) |
From measurement mode, press and hold the ON/OFF button until "CAL" is displayed, then place the probe tip in pH 4.01 or 7.01 buffer. When the reading is stable, "Sto" is displayed and the calibration is saved.
It uses a CR2032 Li-ion battery. Turn the battery cover on the back of the meter counterclockwise to unlock, remove the cover and fit the new battery with the + side facing up.
It appears when the battery has been depleted; the meter then turns off.
The manual states that for best results the meter should be recalibrated periodically, and it should always be calibrated in fresh buffers before use.