DP 300 Mobile Dew Point Instrument Instruction Manual

Instruction manual for the DP 300 mobile dew point instrument: safety, operation, technical data, delivery, calibration, and order information.


DP 300 Mobile Dew Point Instrument Instruction Manual - cover page
Category
Analyzer
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
182 KB
Published
17 November, 2026
Updated
17 November, 2026
MD5 checksum
5ECA782F7C9662763A2E9D1E7792696D

About this manual

The manual describes the DP 300 mobile dew point instrument as a service instrument in a case for measuring pressure, dew point, relative humidity, and temperature in applications such as compressed air plants, granulate dryers, medical gases, and non-corrosive gases. It states that the instrument conforms to DIN EN 61326-1 and DIN EN 61010-1.

The document covers safety notes, dry storage, operation controls, technical data, scope of delivery, service/calibration/adjustment, warranty terms, order data, and humidity reference cells. It also lists standard and high-pressure versions, PC software parameters, measuring chambers, a Teflon cable, a dry container, and a transport case.

What's inside

  • Notes on safety
  • Introduction
  • Measurable gases
  • Container for the dry storage of the instrument
  • Operation DP 300
  • Technical data
  • Scope of delivery
  • Service/calibration/adjustment
  • Order data
  • Humidity Reference Cells

Specifications

Dew point measuring range-80….50 °C td
Temperature measuring range-20….70°C
Relative humidity measuring range0….100 % RH
Standard pressure range-1 up to 50 bar
High-pressure version range-1 up to 350 bar
DisplayOne line, resolution 0,1
Display functionsMin, Max, battery load
Accuracy (-10...50°C td)± 0,5 °C td
Typical accuracy at -40°C td± 2°C td
BatteryInternal rechargeable battery (4 x 1,5 NiMh AAA)
Battery operating timeappr. 15 h operation
Operating temperature-20...70°C measuring gas temperature; 0...40 °C ambient temperature
EMCDIN EN 61326-1
Screw-in threadG ½“ stainless steel
WeightDP 300 - 340 g

Frequently asked questions

What does the DP 300 measure?

It measures pressure, dew point, and atmospheric dew point in different applications.

Which applications are listed for the DP 300?

Listed applications include compressed air plants, granulate dryers, medical gases, and non-corrosive gases such as nitrogen and SF6.

What is the pressure range of the standard DP 300?

The standard version has a pressure range from -1 up to 50 bar.

What should be done before installing the DP 300?

Before installation, briefly bleed the compressed air to remove condensate and particles.

Download PDF (182 KB)