This document is the Thermo Fisher Scientific LTQ Orbitrap series pre-installation requirements guide
Thermo Fisher Scientific published this Revision B (1225821) preinstallation requirements guide for the LTQ Orbitrap Series. It describes site conditions and preparations needed before installation, including laboratory space and floor loading, doorway clearances, main power, temperature and humidity, gases, exhaust, and local safety compliance.
The document includes a Yes/No site checklist, notes for API, MALDI, and ETD configurations, and a declaration of conformity. It also lists Thermo Scientific offices and states that the product is For Research Use Only, not for diagnostic procedures.
| Floor load with API source | 682 kg [1500 Ib] |
|---|---|
| Additional floor load with MALDI source | +50 kg [+110 Ib] |
| Additional floor load with ETD system | +125 kg [+276 lb] |
| Entrance width with API source | at least 89 cm (35 in) |
| Entrance width with MALDI source | at least 110 cm (43 in) |
| Laboratory temperature | 15 to 27 °C (59 to 81 °F) |
| Relative humidity | 50% to 80%, no condensation |
The floor must be certified to meet the load requirements of an LTQ Orbitrap Series system with API source, stated as 682 kg [1500 Ib].
The laboratory must be maintained at a constant temperature between 15 and 27 °C (59 and 81 °F).
Relative humidity must be between 50% and 80%, with no condensation.
The entrance and route from the loading dock must be at least 89 cm (35 in) wide, with additional space at corners.