Thermo Fisher Scientific LTQ Orbitrap Series Instrument Preinstallation Requirements Guide

This document is the Thermo Fisher Scientific LTQ Orbitrap series pre-installation requirements guide


Thermo Fisher Scientific LTQ Orbitrap Series Instrument Preinstallation Requirements Guide - cover page
Brand
Thermo Fisher Scientific
Category
Analyzer
Document type
Installation Guide
Language
English
Pages
100
File format
PDF
File size
3.2 MB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
F8FE86E7951E2CC31157A2FAFD5F422F

About this manual

Thermo Fisher Scientific published this Revision B (1225821) preinstallation requirements guide for the LTQ Orbitrap Series. It describes site conditions and preparations needed before installation, including laboratory space and floor loading, doorway clearances, main power, temperature and humidity, gases, exhaust, and local safety compliance.

The document includes a Yes/No site checklist, notes for API, MALDI, and ETD configurations, and a declaration of conformity. It also lists Thermo Scientific offices and states that the product is For Research Use Only, not for diagnostic procedures.

What's inside

  • Site requirements checklist
  • Floor load requirements
  • Doorway and hallway clearances
  • Main power and electrical outlets
  • Temperature and humidity control
  • Gases and gas regulators
  • Exhaust system and telephone
  • Preinstallation material checklists
  • Declaration of conformity
  • Offices for Thermo Scientific Products

Specifications

Floor load with API source682 kg [1500 Ib]
Additional floor load with MALDI source+50 kg [+110 Ib]
Additional floor load with ETD system+125 kg [+276 lb]
Entrance width with API sourceat least 89 cm (35 in)
Entrance width with MALDI sourceat least 110 cm (43 in)
Laboratory temperature15 to 27 °C (59 to 81 °F)
Relative humidity50% to 80%, no condensation

Frequently asked questions

What floor load must the site support for an LTQ Orbitrap Series system with API source?

The floor must be certified to meet the load requirements of an LTQ Orbitrap Series system with API source, stated as 682 kg [1500 Ib].

What laboratory temperature range is required?

The laboratory must be maintained at a constant temperature between 15 and 27 °C (59 and 81 °F).

What relative humidity is required?

Relative humidity must be between 50% and 80%, with no condensation.

How wide must the entrance be for an LTQ Orbitrap Series system with API source?

The entrance and route from the loading dock must be at least 89 cm (35 in) wide, with additional space at corners.

Download PDF (3.2 MB)