PerkinElmer STA 8000 Thermal Analysis System Installation and Hardware Guide

PerkinElmer STA 8000 Installation and Hardware Guide. This manual introduces how to install and configure PerkinElmer STA 8000 thermal analyzer.


PerkinElmer STA 8000 Thermal Analysis System Installation and Hardware Guide - cover page
Brand
PerkinElmer
Category
Analyzer
Document type
Installation Guide
Language
English
Pages
61
File format
PDF
File size
690 KB
Published
16 October, 2023
Updated
16 October, 2023
MD5 checksum
92B013359C08CBD9780E718200E0DB01

About this manual

This guide provides information on installing the PerkinElmer STA 8000 Thermal Analysis System and on operating and maintaining its hardware. The installation procedure covers preparing the laboratory, unpacking the system, setting up components such as purge gases and cooling supply, connecting the components, configuring the analyzer, and calibrating it.

The document also includes safety and regulatory information, electrical and environmental requirements, instructions for installing multiple analyzers, important specifications, and an STA 8000 hardware section with sample handling and maintenance procedures. A part-number list covers sensors, sample pans, calibration standards, tubing, and other supplies.

What's inside

  • Introduction
  • Pyris Installation
  • Safety and Regulatory Information
  • Prepare the Laboratory
  • Install Multiple Analyzers
  • Install an STA 8000
  • Important Specifications
  • STA 8000 Hardware
  • STA 8000 Maintenance
  • STA 8000 Part Numbers

Specifications

Maximum power consumption450 W
Supply voltage100-240VAC (+5%, -10%)
Supply voltage frequency50–60 Hz
FusesT6.3A (250VAC type only)
Safe temperature range10–40 °C
Dimensions (L x H x D)36 x 20 x 35 cm
Weight16 kg

Frequently asked questions

What is the maximum power consumption of the STA 8000?

The maximum power consumption is 450 W.

What is the safe temperature range for the STA 8000?

The safe temperature range is 10–40 °C.

How is the furnace cleaned when contaminated?

Heat the furnace up to 1400 °C in an air or oxygen atmosphere for approximately 10 minutes.

What should be done if the sensor becomes contaminated?

Remove it from the furnace and clean it gently with a small brush.