PerkinElmer STA 8000 Installation and Hardware Guide. This manual introduces how to install and configure PerkinElmer STA 8000 thermal analyzer.
This guide provides information on installing the PerkinElmer STA 8000 Thermal Analysis System and on operating and maintaining its hardware. The installation procedure covers preparing the laboratory, unpacking the system, setting up components such as purge gases and cooling supply, connecting the components, configuring the analyzer, and calibrating it.
The document also includes safety and regulatory information, electrical and environmental requirements, instructions for installing multiple analyzers, important specifications, and an STA 8000 hardware section with sample handling and maintenance procedures. A part-number list covers sensors, sample pans, calibration standards, tubing, and other supplies.
| Maximum power consumption | 450 W |
|---|---|
| Supply voltage | 100-240VAC (+5%, -10%) |
| Supply voltage frequency | 50–60 Hz |
| Fuses | T6.3A (250VAC type only) |
| Safe temperature range | 10–40 °C |
| Dimensions (L x H x D) | 36 x 20 x 35 cm |
| Weight | 16 kg |
The maximum power consumption is 450 W.
The safe temperature range is 10–40 °C.
Heat the furnace up to 1400 °C in an air or oxygen atmosphere for approximately 10 minutes.
Remove it from the furnace and clean it gently with a small brush.