Hanna Instruments HI 8730 Portable EC TDS Temperature Meter Instruction Manual


Hanna Instruments HI 8730 Portable EC TDS Temperature Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Analyzer
Document type
User Manual
Language
English
Pages
2
File format
PDF
Published
10 July, 2012
Updated
29 September, 2023
MD5 checksum
C0A5B301C656513CDF9450EECB4B9793

About this manual

The HI 8730 is a water-tight portable meter from Hanna Instruments for measuring EC in the 0 to 1990 µS/cm range, TDS from 0 to 1990 ppm, and temperature from 0 to 70°C. It uses the HI 761285 conductivity probe with built-in temperature sensor and has a membrane keyboard for selecting measurement modes.

The manual covers preliminary examination, functional description, specifications, taking measurements, EC/TDS calibration, probe maintenance, accessories and battery replacement. EC/TDS readings use a fixed conversion factor of 0.5, and temperature compensation uses a fixed coefficient of 2%/°C.

What's inside

  • Preliminary Examination
  • General Description
  • Functional Description
  • Specifications
  • Taking Measurements
  • EC/TDS Calibration
  • Probe Maintenance
  • Accessories
  • Battery Replacement
  • EC/TDS Conversion Factor

Specifications

EC Range0 to 1990 µS/cm
TDS Range0 to 1990 ppm
Temperature Range0 to 70°C
TDS Conversion Factor0.5
Battery Type and Life1x 9V alkaline, approximately 250 hours
Dimensions164 x 76 x 45 mm
Weight750 g (8.8 oz.)
Environment0 to 50°C (32 to 122°F); RH max 100%
ProbeHI 761285 with built-in temperature sensor and 1 m cable

Frequently asked questions

What is the EC/TDS conversion factor on the HI 8730?

The meter uses a fixed conversion factor of 0.5, so 1 µS/cm equals 0.5 ppm of TDS.

How is temperature compensation applied?

EC measurements are automatically compensated for temperature changes with a fixed coefficient of 2%/°C.

Which battery does the HI 8730 use?

It is powered by a 9V alkaline battery, with a low-battery indication on the display.

How many calibration points are used for EC or TDS?

The meter is calibrated at one point for EC or TDS, and the other range is automatically calibrated.

Download PDF

More Hanna Instruments Manuals


View all Hanna Instruments manuals →