The BANTE PHscan 30F is a pen-style pH meter with an LCD display, automatic temperature compensation, support for 1 to 3 point pH calibration, and manual or automatic data hold. It can use USA or NIST pH buffer standards and shuts off automatically after 8 minutes without key operation.

The manual explains power on and off, setup menu options, pH and temperature calibration, pH measurement, electrode cleaning and maintenance, electrode and battery replacement, troubleshooting codes, technical specifications, and warranty terms. This manual is written in Chinese.

What's inside

  • Key functions and display icons
  • Preparation and pH buffer solution
  • Power on/off and setup menu
  • pH calibration
  • Temperature calibration
  • pH measurement and data hold
  • Electrode cleaning and maintenance
  • Troubleshooting
  • Technical specifications
  • Warranty

Specifications

pH measurement range-1.00 to 15.00 pH
pH measurement accuracy±0.01 pH
pH calibration points1 to 3 points
pH buffer standardsUSA (pH 4.01/7.00/10.01) or NIST (pH 4.01/6.86/9.18)
Temperature compensation range0 to 60 °C (32 to 140 °F)
Temperature compensationAutomatic
Temperature measurement range0 to 60 °C (32 to 140 °F)
Temperature measurement accuracy±1 °C
Temperature calibration rangeMeasured value ±10 °C
Data holdManual or automatic endpoint lock
Power off modeManual or automatic after 8 minutes without key operation
Sensor typeE-PHSCAN-FT
Operating temperature0 to 60 °C (32 to 140 °F)
Power supply2 x 1.5 V AAA batteries
Dimensions185 (length) x 40 (diameter) mm
Weight100 g

Frequently asked questions

How do I turn off the meter?

Hold the MEAS key for 3 seconds.

Does the meter shut off automatically?

Yes, after 8 minutes without key operation.

How many pH calibration points are supported?

1 to 3 points.

Which pH buffer standards can be used?

USA (pH 4.01, 7.00, 10.01) or NIST (pH 4.01, 6.86, 9.18).

What is the warranty period?

One year from the date of shipment; the sensor and calibration solutions are not covered.