This document introduces an industrial electrode that is suitable for detecting the pH and temperature of water solutions. The electrode has a wide measurement range and can operate within the pH range of 0 to 14. It is also equipped with a temperature sensor for measuring within the temperature range of 0 to 60°C. The electrode has high pressure resistance and weather resistance, making it suitable for continuous measurement in industrial processes. The document also provides information on various installation methods for the electrode, as well as maintenance methods and optional accessories.
The Bante IE-20T is an industrial pH electrode designed for continuous measurement of pH and temperature in water solutions. It features a PPS/PC housing, Ag/AgCl reference type, and an NTC 10 KΩ temperature sensor with a range of 0 to 60°C. This manual is written in Chinese.
| Measurement Range | 0 to 14 pH |
|---|---|
| Temperature Sensor Type | NTC 10 KΩ |
| Operating Temperature | 0 to 60°C (32 to 140°F) |
| Maximum Pressure Resistance | 0.6 MPa (87 PSIG) |
| Housing Material | PPS/PC |
| Reference Type | Ag/AgCl |
| Electrode Diameter | 29.5 mm (1.16 in.) |
| Thread Type | 3/4 inch NPT |
| Standard Cable Length | 5 m (16.4 ft.) |
| Maximum Transmission Distance | 30 m (98.4 ft.) |
Add electrode immersion liquid inside the transparent protective cap and install it on the pH electrode.
No, do not use distilled or deionized water as it will exhaust the hydration layer of the pH sensitive membrane and cause permanent damage.
Soak the electrode in flexible detergent for 30 minutes, rinse with water, and immerse in electrode immersion solution for about 1 hour.
Soak the electrode in 0.1M hydrochloric acid for 10 minutes, rinse with water, and immerse in electrode immersion solution for at least 6 hours.