This document describes the usage and maintenance requirements of a series of electrodes. These electrodes are suitable for measuring the acidity or alkalinity (pH) of solutions and have features such as high sensitivity and ease of use.
The Bante pH Electrode I is designed to detect the acidity or alkalinity (pH) of solutions. Before use, remove the transparent protective cap at the bottom of the electrode. If small bubbles exist inside the pH sensitive membrane, gently shake the electrode downwards to remove them. Calibration should be performed according to the instrument's user manual before measurement. This manual is written in Chinese.
No, do not use distilled or deionized water as it will exhaust the hydration layer of the pH sensitive membrane and cause permanent damage.
Soak the electrode in flexible detergent or ethanol for 15 minutes, then rinse with distilled water and immerse it in the electrode soaking solution.
If the response is slow, immerse the electrode in pH 4.01 standard buffer for 30 minutes; if still unresponsive, soak it in 0.1M HCl for 10 minutes followed by 0.1M NaOH and finally 3M KCl.