Product manual for the Aosong AO-09 oxygen sensor, covering specifications, dimensions, installation, use, and precautions.
The AO-09 is an Aosong Electronics oxygen sensor for medical oxygen analyzers, with a measurement range of 0 to 100 Vol.% O2, linear output, temperature compensation, and an expected operating life stated for 20 C and 40 C.
The manual lists technical specifications, product dimensions, installation and use requirements, and warnings and precautions for the sensor.
| Operating principle | Partial pressure electrochemical |
|---|---|
| Output | 9 mV to 13 mV in 210 mBar O2 |
| Measurement range | 0 - 100 Vol.% O2 |
| Housing material | White ABS |
| Weight | 36 g (nominal) |
| Operating pressure range | 0.5 Bar to 2.0 Bar |
| Operating humidity range | 0% to 99% RH noncondensing |
| Expected operating life at 20 C | 1.5 x 106%O2 hours |
| Standard warranty | 13 months from date of despatch |
It is intended as an oxygen-sensing component of an oxygen analyzer that measures oxygen concentration in breathing gas mixtures for applications such as oxygen concentrators, medical ventilators, anaesthesia equipment, and incubators.
Allow at least 15 minutes to stabilize in the instrument before calibration.
No, the sensor is not suitable for sterilisation by steam or exposure to chemicals such as ethylene oxide or hydrogen peroxide.