AS200 series manual, for Aosong AS200 series gas mass flow controller
This manual provides detailed instructions for installing, operating, and maintaining AOSONG AS200 series gas mass flow controllers. It covers safety precautions, mechanical and electrical installation procedures, technical specifications, and signal compatibility. The device supports digital, 1–5V analog, and 4–20mA analog signals, operates on single +15 to +24VDC power, and features high precision and low leakage. It is designed for use in semiconductor, chemical, pharmaceutical, and other industrial applications.
The manual includes warnings for handling hazardous gases, proper installation orientation, leak testing, and calibration procedures. It also details connection methods for D-type 9-pin, RS485, and EtherCAT interfaces, along with mechanical fitting guidelines for VCR and double flange connectors. This manual is written in Chinese.
| Accuracy | ±1.0%S.P.(≥35%F.S.) ±0.35%F.S.(<35%F.S.) |
|---|---|
| Repeatability | ±0.2%F.S. |
| Leak Rate | 1×10-10Pa·m3/secHe |
| Operating Voltage | +15~+24VDC |
| Signal Output | Digital, 1–5V analog, 4–20mA analog |
| Max Working Pressure | 0.45MPa |
| Temperature Coefficient | Zero drift: ≤±0.05%F.S./℃; Adjustment: ≤±0.1%F.S./℃ (≤30SLM), ≤±0.2%F.S./℃ (>30SLM) |
| Response Time | 1sec |
| Sealing Material | Fluororubber |
| Weight | 约0.95Kg |
It uses a single +15 to +24VDC power supply.
It supports digital signals, 1–5V analog signals, and 4–20mA analog signals.
The leakage rate is 1×10-10Pa·m3/secHe.
The maximum working pressure is 0.45MPa.