Start here

Aosong AO-09 Oxygen Sensor Product Manual
The AO-09 is an Aosong Electronics oxygen sensor for medical oxygen analyzers, with a measurement range of 0 to 100 Vol.% O2, linear output, temperature compensation, and an expected operating life stated for 20 C and 40 C.
The manual lists technical specifications, product dimensions, installation and use requirements, and warnings and precautions for the sensor.
Aosong AO-09 specifications
| Operating principle | Partial pressure electrochemical |
|---|---|
| Output | 9 mV to 13 mV in 210 mBar O2 |
| Measurement range | 0 - 100 Vol.% O2 |
| Housing material | White ABS |
| Weight | 36 g (nominal) |
| Operating pressure range | 0.5 Bar to 2.0 Bar |
| Operating humidity range | 0% to 99% RH noncondensing |
| Expected operating life at 20 C | 1.5 x 106%O2 hours |
| Standard warranty | 13 months from date of despatch |
All Aosong AO-09 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the AO-09 oxygen sensor used for?
It is intended as an oxygen-sensing component of an oxygen analyzer that measures oxygen concentration in breathing gas mixtures for applications such as oxygen concentrators, medical ventilators, anaesthesia equipment, and incubators.
How long should the sensor stabilize before calibration?
Allow at least 15 minutes to stabilize in the instrument before calibration.
What are the intended applications for the AO-09 sensor?
The sensors are intended as oxygen-sensing components for medical oxygen analyzers, control devices of oxygen concentrators, medical ventilators, anaesthesia equipment, and incubators.
Can the AO-09 sensor be sterilized by steam?
No, the sensor is not suitable for sterilization by steam or exposure to chemicals such as ethylene oxide or hydrogen peroxide.