Training manual for operating the Agilent PrepStar SD-1 preparative HPLC, including method setup, data acquisition, and column use.
The Agilent PrepStar SD-1 is a dual-pump preparative HPLC with analytical, semi-prep, and prep modes. It uses three C18 column sizes, two manual injectors with 500 ul and 10 ml sample loops, and requires appropriate guard columns.
The manual covers operating instructions, direct device control, new method setup, data acquisition, and a test-mix standard experiment with operating conditions.
| Pump configuration | Dual pump (binary) |
|---|---|
| Pump capacity | 3.2 L/min |
| Maximum pressure | 375 psi |
| Analytical column | 4.6 x 50 mm, 1.8 um |
| Semi-prep column | 9.4 x 250 mm, 5 um |
| Prep column | 21.2 x 250 mm, 7 um |
| Flow rates | 1, 4.2, and 21.0 ml/min |
| Sample loads | 2-10, 10-50, and 50-200 mg |
| Injection volumes | 1-20 ul, 100-400 ul, 500 ul-5 ml |
| Analytical/semi-prep injector loop | 500 ul |
| Prep injector loop | 10 ml |
The bottom injector, which is equipped with a 10 ml sample loop.
0.000 ml/min.
Methods are in C:\Chem32\1\METHODS and data files are in C:\Chem32\1\DATA.
Use the buttons on the main instrument page.