Agilent 34410A/11A 6½ digit multimeter user guide, introduces the basic function, usage, precautions and other information of the product
This user guide covers the Agilent 34410A and 34411A 6½ digit digital multimeters. It describes DC and AC (true RMS) voltage and current measurements, 2-wire and 4-wire resistance, continuity, diode, frequency, period, capacitance, and temperature measurements, along with ranging, math functions, data logging, and instrument state storage.
The guide also documents safety information, protection limits, IEC Measurement Category II overvoltage protection, the included Agilent 34138A test lead set, and the GPIB (IEEE-488), USB, LAN, web interface, and SCPI compatibility. The 34411A differs from the 34410A in maximum reading speed and buffer capacity. This manual is written in Chinese.
| Display resolution | 6½ digits |
|---|---|
| 34410A maximum reading rate | 10,000 readings per second |
| 34411A maximum reading rate | 50,000 readings per second |
| 34410A buffer capacity | 50,000 readings |
| 34411A buffer capacity | 1 million readings |
| HI to LO protection limit | 1000 VDC or 750 VAC |
| Maximum HI input voltage to ground | 1500 Vpk |
| LO to ground protection limit | 500 Vpk |
| Current input protection limit | 3 A (rms) |
| Sense terminal protection limit | 200 Vpk |
| IEC Measurement Category II maximum line voltage | 300 VAC |
| Transient overvoltage withstand | 2500 Vpk |
| Included test lead set | Agilent 34138A |
| Test lead rating | 1000V, 15A |
| Remote interfaces | GPIB (IEEE-488), USB, and LAN |
| LXI standard class | LXI Class C |
| Safety standard | IEC 61010-1:2001 / EN 61010-1:2001 |
They perform DC and AC (true RMS) voltage and current, 2-wire and 4-wire resistance, continuity, diode, frequency, period, capacitance, and temperature measurements.
The 34411A reads up to 50,000 readings per second and stores up to 1 million readings in its buffer, while the 34410A reads up to 10,000 readings per second and stores up to 50,000 readings.
The HI to LO protection limit is 1000 VDC or 750 VAC, which is also the maximum voltage measurement limit.
The multimeter provides three standard remote interfaces: GPIB (IEEE-488), USB, and LAN, plus a web interface for direct browser access.