The Agilent 8860 Gas Chromatograph Data Sheet provides specifications, features, and performance details for the 8860 GC system.
About this manual
The Agilent 8860 Gas Chromatograph features a 10‑inch color touchscreen, a browser interface for remote access, and full electronic pneumatics control (EPC) with optional electronic pneumatics regulation (EPR). It supports up to two inlets and three detectors, offering precise pressure and flow control, advanced diagnostics, and method editing without a separate data system.
The system includes a column oven with a temperature range up to 425 °C, rapid ramp rates, and multiple safety and regulatory certifications. Power requirements vary by voltage, and the instrument is designed for operation in ambient temperatures from 15 °C to 35 °C with robust environmental tolerances.
What's inside
- Data Sheet
- Full electronic pneumatics control (EPC)
- Power requirements
- Column oven
- Inlets
- Detectors
- Other EPC Devices
- Data communications
- Safety and regulatory certifications
- Other specifications
Specifications
| Power consumption | 1,500 W at 100 V; 2,250 W at other voltages |
|---|
| Oven temperature range | 8 °C above ambient to 425 °C |
|---|
| Maximum oven ramp rate | 75 °C/min |
|---|
| EPC pressure setpoint precision | 0.01 psi (0.069 kPa) |
|---|
| Maximum inlet pressure range | 0–100 psi (0–689.47 kPa) |
|---|
| Maximum detector temperature (FID) | 425 °C |
|---|
| Dimensions (H × W × D) | 49 cm × 58 cm (68 cm with side detector) × 54 cm |
|---|
| Weight | 51 kg |
|---|
| Ambient operating temperature | 15 °C to 35 °C |
Frequently asked questions
What is the power consumption of the Agilent 8860 GC?
The system consumes 1,500 W at 100 V and 2,250 W at all other supported voltages.
How many inlets can be installed on the 8860?
Up to two inlets may be installed.
What is the maximum oven temperature?
The column oven can reach up to 425 °C.