Datasheet for Agilent 1260 Infinity II Refractive Index Detector, designed for non-UV absorbing compounds with stable baselines and fast equilibration.
The Agilent 1260 Infinity II Refractive Index Detector is designed for routine analysis of compounds lacking UV chromophores, such as carbohydrates and pharmaceutical excipients. It features rapid equilibration, stable baselines, and electronic temperature control for consistent results. It supports polymer characterization via GPC and includes automatic lamp purging, solvent recycling, and extensive diagnostics for maintenance and error tracking. This manual is written in Chinese.
| Detection Type | Refractive Index |
|---|---|
| Refractive Index Range | 1.00 – 1.75 RIU |
| Measurement Range | ±60 0·10-6 RIU |
| Optics Temperature Control | 5 °C above ambient to 55 °C |
| Weight | 15 kg (33 lbs) |
| Dimensions | 180 x 396 x 436 mm |
| Line Voltage | 100 – 240 V~ |
| Power Consumption | 70 W / 80 VA |
| Ambient Operating Temperature | 4–55 °C |
| Maximum Data Rate | 74 Hz |
It is suitable for compounds without UV chromophores, such as carbohydrates, triglycerides, organic acids, and pharmaceutical excipients.
The measurement range is 1.00 to 1.75 RIU, calibrated.
The maximum data rate is 74 Hz, ensuring even the narrowest peaks can be detected and quantified.
The ambient operating temperature range is 4–55 °C.