This document outlines qualification procedures for Agilent 1100 Series GPC-SEC systems, covering software validation, system verification, and performance…
This document addresses the need for comprehensive qualification procedures for GPC-SEC (Gel Permeation Chromatography-Size Exclusion Chromatography) systems in laboratories regulated by GLP or ISO 9000 standards. It highlights that standard HPLC qualification tests are often insufficient for GPC-SEC, necessitating specific procedures to ensure the accuracy and reliability of molecular weight data.
| System verification limit | 0.5% |
|---|---|
| Resolution limit | > 1.7 |
| Efficiency limit | > 6cm |
The procedures are compliant with ASTM D 5296, ISO 13885, and DIN 55672.
They are designed to enable a fast entry into and a successful life in the complex world of GPC-SEC.
It verifies that the ChemStation GPC data analysis software is calculating properly using a known theoretical Schulz-Flory distribution.