Beckman Coulter Type 45 Ti Rotor Manual

This document describes the safety and reliability of the Beckman Coulter Type 45 Ti rotor.


Beckman Coulter Type 45 Ti Rotor Manual - cover page
Brand
Beckman Coulter
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
28
File format
PDF
File size
1.1 MB
Published
29 November, 2023
Updated
29 November, 2023
MD5 checksum
5D8C0B11EBE842CF880B0F95DA2FB688

About this manual

The Type 45 Ti is a titanium fixed-angle rotor rated for 45,000 rpm and designed to centrifuge up to six tubes at a 24-degree angle. It is used in Beckman Coulter Class H, R, and S preparative ultracentrifuges for applications such as pelleting small particles and purifying viruses and subcellular organelles.

The rotor has a nominal capacity of 564 mL and a largest tube capacity of 94 mL. The manual covers safety notices, specifications, preparation, operation, removal, and tube and bottle compatibility.

What's inside

  • Safety Notice
  • Type 45 Ti Rotor Specifications
  • Description
  • Preparation and Use
  • Prerun Safety Checks
  • Rotor Preparation
  • Operation
  • Removal and Sample Recovery
  • Tubes and Bottles

Specifications

Maximum speed45 000 rpm
Density rating at maximum speed1.2 g/mL
Relative centrifugal field at rmax235 000 × g
Relative centrifugal field at rav158 000 × g
Relative centrifugal field at rmin81 400 × g
k factor at maximum speed133
Number of tube cavities6
Nominal tube dimensions (largest tube)38 × 102 mm
Nominal tube capacity (largest tube)94 mL
Nominal rotor capacity564 mL
Approximate acceleration time9 min
Approximate deceleration time9 1/2 min
Weight of fully loaded rotor12 kg (26 lb)
Rotor materialtitanium
Tube angle24 degrees

Frequently asked questions

What is the maximum speed of the Type 45 Ti rotor?

The maximum speed is 45,000 rpm.

How many tubes can be centrifuged per run?

Two, three, four, or six tubes can be centrifuged per run if they are arranged symmetrically in the rotor.

What is the rotor made of?

The rotor is made of titanium.

What should be done if leakage is found during disassembly?

Assume that some fluid escaped the rotor and apply appropriate decontamination procedures to the centrifuge and accessories.