This product is a digital ground tester with functions such as continuity test, R ground test, R loop resistance measurement, R insulation measurement, AC voltage measurement and RCD test.
This is the Chinese user manual for the MX 435D digital ground tester, a portable instrument powered by a rechargeable battery or two 9 V alkaline batteries. The manual explains safety precautions for indoor use in CAT III installations, the symbols printed on the front panel, warranty and cleaning, and describes the controls: the ON/OFF switch, LCD display, E/S/H ground terminals, RCD/VAC/AAC position, continuity position, MΩ insulation terminal, voltage indicator LED, TEST button and 6-position rotary switch.
The second half describes the measurement functions — continuity with buzzer, ground resistance, loop resistance, insulation at 500 VDC, AC voltage, current with a current clamp and RCD testing — followed by technical specifications such as continuity range 0.10 - 20.00 Ω, ground resistance ranges 0.15 - 20.00 Ω and 15 - 2000 Ω, test frequency 1024 Hz, protection 480 Vrms and RCD test currents of 30, 100, 300, 500 and 650 mA. This manual is written in Chinese.
| Continuity range | 0.10 - 20.00 Ω |
|---|---|
| Continuity resolution | 0.01 Ω |
| Continuity measuring current | > 200 mADC |
| Continuity open circuit voltage | 4 V ≤ V ≤ 24 VDC |
| Buzzer threshold | 2 Ω ± 0.15 Ω |
| Ground resistance ranges | 0.15 - 20.00 Ω; 15 - 2000 Ω |
| Test voltage frequency | 1024 Hz |
| Protection | 480 Vrms |
| RCD test currents | 30, 100, 300, 500, 650 mA |
| Rechargeable battery | 8.4 V Ni-MH, 750 mAh |
| Alternative battery supply | 2 x 9 V alkaline, type 6LF22 |
| Charging time | 2 hours for a full charge |
Turn the rotary switch to "I" to switch the instrument on, and to switch it off, unplug all leads and turn the switch to "O".
It switches off automatically after ten minutes without use; the five RCD range lights roll to warn you, and the TEST button can delay the shutdown.
It means the voltage is too low, so the battery must be replaced or recharged.
In continuity test mode the buzzer sounds when the measured resistance is less than 2 Ω.