Quick Reference Guide for Thermo Scientific Dionex IC Pure Ultrapure Water System.
This document serves as a quick reference guide for the Thermo Fisher Scientific Dionex IC Pure Ultrapure Water System. It provides essential information regarding the replacement schedules for various consumables, such as the ultrapure cartridge, final filter, UV lamp, and pretreatment cartridges, along with their estimated lifetimes and replacement triggers. The guide also outlines the system's display functionality and key operational modes, including Nonstop, Analyser, and Interval modes.
| Replacement interval for 5 µm filter with carbon cartridge | 3-4 months |
|---|---|
| Replacement interval for 10" 5 µm filter with hardness stabilizer cartridge | 6-12 months |
| Replacement interval for ultrapure cartridge | Up to 12 months depending on feed water quality |
| Replacement interval for 0.2 μm final filter | Up to 12 months |
| Replacement interval for Ultrafilter | Up to 2 years |
| Replacement interval for UV lamp | Up to 2 years |
| Replacement interval for Vent filter | Up to 12 months |
| Automatic switching to interval mode duration | 2 hours of non-use |
Replace when purity drops below acceptable levels, the 'Lim. val. pure w.' display appears, or a 'Change cartridge' message is shown.
The unit automatically switches into interval mode after 2 hours of non-use.
It indicates that the feed water purity limit is out of range, requiring a check of the feed water supply.