Thermo Fisher Scientific Dionex IC Pure Ultrapure Water System Quick Reference Guide

Quick Reference Guide for Thermo Scientific Dionex IC Pure Ultrapure Water System.


Thermo Fisher Scientific Dionex IC Pure Ultrapure Water System Quick Reference Guide - cover page
Brand
Thermo Fisher Scientific
Category
Other
Document type
Quick Start Guide
Language
English
Pages
2
File format
PDF
File size
438 KB
Published
19 November, 2023
Updated
19 November, 2023
MD5 checksum
93C6C42D5BD986A8E34DE10D24DB7D0B

About this manual

This document serves as a quick reference guide for the Thermo Fisher Scientific Dionex IC Pure Ultrapure Water System. It provides essential information regarding the replacement schedules for various consumables, such as the ultrapure cartridge, final filter, UV lamp, and pretreatment cartridges, along with their estimated lifetimes and replacement triggers. The guide also outlines the system's display functionality and key operational modes, including Nonstop, Analyser, and Interval modes.

What's inside

  • Consumable Part Replacement
  • External Pretreatment
  • Display Functionality
  • Operational Modes
  • Maintenance
  • Common Alerts

Specifications

Replacement interval for 5 µm filter with carbon cartridge3-4 months
Replacement interval for 10" 5 µm filter with hardness stabilizer cartridge6-12 months
Replacement interval for ultrapure cartridgeUp to 12 months depending on feed water quality
Replacement interval for 0.2 μm final filterUp to 12 months
Replacement interval for UltrafilterUp to 2 years
Replacement interval for UV lampUp to 2 years
Replacement interval for Vent filterUp to 12 months
Automatic switching to interval mode duration2 hours of non-use

Frequently asked questions

When should the ultrapure cartridge be replaced?

Replace when purity drops below acceptable levels, the 'Lim. val. pure w.' display appears, or a 'Change cartridge' message is shown.

How does the system switch to Interval mode automatically?

The unit automatically switches into interval mode after 2 hours of non-use.

What does the 'Lim. Val. Feed' alert mean?

It indicates that the feed water purity limit is out of range, requiring a check of the feed water supply.