This document is the disinfection kit manual for the AIRVO™ 2. It contains instructions on how to disinfect the AIRVO™ 2 using the kit.
The Disinfection Kit 900PT600 is used to achieve a high level of disinfection of the AIRVO 2/myAIRVO 2 outlet elbow. It describes recommended reprocessing between patients, including cleaning, high-level disinfection with the supplied disinfection tube and filter, and correct storage.
The manual covers warnings, package contents, cleaning of the outlet elbow and external surfaces, the disinfection cycle and its operating conditions, storage, disinfection history, and symbol definitions. This manual is written in Spanish.
| Kit model | 900PT600 |
|---|---|
| Compatible device | AIRVO 2/myAIRVO 2 humidifier |
| Disinfection temperature | At least 87 °C (189 °F) |
| Minimum disinfection time | 30 minutes |
| Disinfection cycle duration | 55 minutes |
| Calibration time | Approximately 2 minutes |
| Ambient temperature | 18 - 30 °C (64 - 86 °F) |
It is used to achieve a high level of disinfection of the AIRVO 2/myAIRVO 2 outlet elbow.
It must only be used with the AIRVO 2/myAIRVO 2 humidifier.
The cycle is performed at an ambient temperature of 18 - 30 °C (64 - 86 °F).
After calibration, the disinfection cycle is carried out for 55 minutes.