Fisher & Paykel 900PT600 Manuals

ManualLib has 2 documents for the Fisher & Paykel 900PT600 (Professional medical equipment): 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
900PT600
Documents
2
Languages
ES · EN
Last added
2023-11-17

Start here

Fisher & Paykel 900PT600 AIRVO 2 Disinfection Kit Manual

Fisher & Paykel 900PT600 AIRVO 2 Disinfection Kit Manual

User Manual 148 pages · 14.4 MB · EN

This manual is intended for healthcare professionals and describes reprocessing procedures for the Fisher & Paykel 900PT600 Disinfection Kit used with the AIRVO 2 / myAIRVO 2 humidifier.

It covers cleaning, high-level disinfection using the supplied disinfection tube and filter, and correct storage between patients.

Read online

Fisher & Paykel 900PT600 specifications

Kit model900PT600
Compatible deviceAIRVO 2/myAIRVO 2 humidifier
Disinfection temperatureAt least 87 °C (189 °F)
Minimum disinfection time30 minutes
Disinfection cycle duration55 minutes
Calibration timeApproximately 2 minutes
Ambient temperature18 - 30 °C (64 - 86 °F)

Source: Fisher & Paykel AIRVO 2 Disinfection Kit (900PT600) Manual

All Fisher & Paykel 900PT600 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the Disinfection Kit 900PT600 used for?

It is used to achieve a high level of disinfection of the AIRVO 2/myAIRVO 2 outlet elbow.

From: Fisher & Paykel AIRVO 2 Disinfection Kit (900PT600) Manual

Which device can the kit be used with?

It must only be used with the AIRVO 2/myAIRVO 2 humidifier.

From: Fisher & Paykel AIRVO 2 Disinfection Kit (900PT600) Manual

What is the 900PT600 Disinfection Kit used for?

It is used to achieve high-level disinfection of the AIRVO 2 / myAIRVO 2 humidifier outlet elbow.

From: Fisher & Paykel 900PT600 AIRVO 2 Disinfection Kit Manual

Which chamber port should not be cleaned?

Do not clean the left-hand chamber port, because this can damage the non-return valve located at that port.

From: Fisher & Paykel 900PT600 AIRVO 2 Disinfection Kit Manual