The AT3600 is a versatile transformer tester with a constant current drive function that can test a variety of transformers and inductor components.
This document explains how to test transformers, inductors, and cores under constant AC current conditions using the Voltech AT3600 transformer tester. It covers the built-in programmable constant current drive, use of a current-sense resistor and step-down transformer, and integration with other AT3600 tests.
It also describes methods for testing above 2 A, including multiple drive turns, the AC Interface accessory, and an external AC source, plus a worked example, programming steps, and safety cautions.
| Maximum generator current | 2 A |
|---|---|
| AC Interface current | up to 10 A |
| Internal generator VA | 500 VA |
| Maximum supply voltage | 270 V |
| Current-sense resistor voltage range | 10 mV to 1 V |
| Current-sense resistor maximum value | 1 ohm |
Use more than one turn of the drive signal through the part under test, use the AC Interface accessory, or use an external AC source.
The voltage produced should be in the range 10 mV to 1 V.
Yes, other tests such as R can be added using extra nodes and an extra dummy winding.
Yes, an external AC source may be used to increase the VA rating of the AT3600's generator.