Voltech AT Series Transformer Testers AT Calibration System User Manual (AT3600 + ATi)
This manual guides users through calibrating Voltech AT series testers such as the ATi and AT3600 using a calibration fixture, PC software, and a precision digital voltmeter. It covers setup, calibration procedures, verification, troubleshooting, and safety precautions. The system supports both manual and automatic calibration, with software generating calibration certificates. Safety warnings emphasize high-voltage risks and require qualified personnel. Calibration ensures measurement accuracy by adjusting internal constants against international standards.
The manual includes detailed sections on installation, calibration steps, verification, stored data, warranty, specifications, and troubleshooting for common issues like communication errors, self-test failures, and certificate printing problems.
Voltech AT3600 specifications
| VA Rating | 2KVA or less |
|---|---|
| Test Functions | Over 30 test functions |
| Connection Matrix | 20-node |
| Line Input | 90 to 265V ac, 48Hz to 65Hz |
| Operating Temperature | 0° to 40°C |
| Storage Temperature | -40° to +70°C |
| Weight | approx. 23kg |
| Basic Accuracy | 0.05% |
All Voltech AT3600 documents by type
How to use the product for one specific task, with worked examples.
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the AT3600 used for?
It is a transformer and wound component tester for 100% functional and safety testing.
How many test functions does the AT3600 offer?
It offers over 30 test functions.
How can the AT3600 test constant current above 2 A?
Use more than one turn of the drive signal through the part under test, use the AC Interface accessory, or use an external AC source.
What voltage should be produced across the current-sense resistor?
The voltage produced should be in the range 10 mV to 1 V.
What is required to calibrate the AT series testers?
A calibration fixture, a PC running the Voltech calibration software, a printer, and a precision digital voltmeter (not supplied).
From: Voltech AT Series Transformer Testers AT Calibration System User Manual (AT3600 + ATi)
What is the recommended calibration interval?
Voltech recommends a calibration interval of one year.