This manual provides installation and operating instructions for Thermo Scientific Forma Series Ultra-Low Temperature Freezers.
This document serves as the installation and operation manual for the Thermo Fisher Scientific 315673H31 ultra-low temperature freezer. It covers essential safety precautions, unpacking procedures, and general recommendations for usage, including initial loading and operating standards. The guide details the electrical specifications and installation requirements, such as proper location, leveling, and wiring connections to ensure safe and efficient operation. This manual is written in Spanish.
| Ambient Temperature Range | 15 to 32°C |
|---|---|
| Maximum Altitude | 2.000 m |
| Maximum Relative Humidity | 60% |
| Voltage Type A | 115V 60Hz 16A |
| Voltage Type D | 208/230V 60Hz 12A |
| Voltage Type V | 230V 50Hz 12A |
You must let the freezer run at the desired temperature for at least 12 hours before loading it.
You must leave a free space of at least 20 cm on the sides and top.
No, this unit is not a rapid freezing device.