Keysight U1401B Handheld Multi-Function Calibrator/Meter Quick Start Guide
The Keysight U1401B Quick Start Guide introduces the handheld multi-function calibrator/meter, its safety information, included accessories, and basic operation. It covers recharging and charging the battery, keypad functions and features, input/output terminals with overload protection, and voltage measurement steps.
The guide also describes the tilt stand and notes that some product specifications can degrade in ambient electromagnetic fields and noise, with self-recovery when the source is removed or the instrument and cabling are protected. It explains selecting AC, DC, or AC+DC, data hold, peak hold, refresh hold, dynamic recording, dual display, frequency measurement, and square-wave output adjustment.
| Measurement category | CAT II 150 V, Pollution Degree 2 |
|---|---|
| Output terminal overload protection | DC 30 V |
| AC/DC voltage input overload protection | 250 Vrms |
| AC/DC current input fuse | 250 V/630 mA, fast-acting |
| Rechargeable battery pack | 1.2 V NiMH AA × 8 |
| AC adapter | 24 V AC |
Set the slide switch at the CHARGE position after connecting the AC adapter; red indicates charging and green indicates fully charged.
Included items are a protective holster, rechargeable battery pack, power cord and AC power adapter, silicone test leads and alligator clips, a yellow test lead for mA simulation, certificate of calibration, and printed Quick Start Guide.
Set the rotary switch to V or mV, select DC, connect the red (+) and black (–) test leads to the INPUT terminals, then probe the test points and read the display.
The AC/DC current measurement input is protected by a 250 V/630 mA fast-acting fuse.