The 6250/6250C is a portable multi-function measuring instrument used in the laboratory to measure pH, temperature, oxidation-reduction potential and direct-reading ion concentration.
This operation manual covers the JENCO 6250/6250C microcomputer-based portable meter for pH, mV, ion concentration, and temperature measurement. It explains the buttons and display indicators, connector layout, battery replacement and power on/off, then gives step-by-step procedures for pH calibration and pH, temperature, mV, and ion measurement.
The manual also describes RS232C interface operation for connecting the meter to a computer, lists error indications, and provides product specifications and warranty information. MODEL 6250 uses 4.00, 7.00, and 10.01 pH calibration solutions; MODEL 6250C uses 4.00, 6.86, and 9.18 pH solutions. This manual is written in Chinese.
| pH range | -2.000 to 16.000 pH |
|---|---|
| pH accuracy | +/- 0.1 % +/- 1 digit |
| pH resolution | 0.01 pH or 0.001 pH, selectable |
| mV range | -1999.9 to 1999.9 mV |
| mV resolution | 1 mV or 0.1 mV, selectable |
| Temperature range | -5.0 to 125.0 °C |
| Ion range | 0.00 to 119.8 conc. |
| Dimensions | 241 mm / 83 mm / 72 mm (L/W/H) |
MODEL 6250 uses 4.00 pH, 7.00 pH, and 10.01 pH standard calibration solutions.
MODEL 6250C uses 4.00 pH, 6.86 pH, and 9.18 pH standard calibration solutions.
It uses six AAA batteries or a UL-approved AC adapter rated 9 VDC, 100 mA.
The pH range is -2.000 to 16.000 pH.