Operation manual for the Jenco Model 60 portable digital pH meter, covering calibration, measurement procedures, electrode care, and specifications.
The document serves as an operation manual for the Jenco Model 60 Portable Digital pH Meter. It provides instructions on initial inspection, general introduction to the instrument's capabilities regarding internal batteries or AC adapter usage, and detailed steps for dual-point calibration using buffers at pH 7.00 and 4.01.
| Range | 0 to 14.00 pH |
|---|---|
| Accuracy | ±0.01 digit |
| Temperature Compensation | Manual 0 to 100 °C |
| Input Impedance | >10^12 Ohms |
| Display | 0.5 inch high LCD |
| Power Source | One 9V battery or AC adapter |
| Battery Life | 1400 hours |
| Dimensions | 150 mm L x 85 mm W x 38 mm H |
Connect the electrode, rinse it with distilled water, immerse in pH buffer 7.00 and adjust calibration control, then repeat with pH buffer 4.01 to adjust the slope control.
Replace the internal 9 volt battery when the BAT LOW indicator is ON.
Rinse the electrode with distilled water after each measurement and soak it in buffer 4.01 and KCL when not being used.
The accuracy is ±0.01 digit.