Installation and Operation Manual for Ultra-Low Temperature Freezers RevcoTM RDE Series, FormaTM FDE Series, HERAfreezeTM HDE Series and Thermo ScientificTM TDE Series.
Revco RDE, Forma FDE, HERAfreeze HDE and Thermo Scientific TDE ultra-low temperature freezers are covered by this installation and operation manual. It describes safety considerations, unpacking, packing list contents, general recommendations, temperature monitoring, general usage, initial loading, battery door operation, operating standards, electrical specifications, installation, operation, optional backup systems, CO2 and LN2 precautions, optional chart recorders, maintenance, troubleshooting, error codes, warranty and contact information.
The manual lists applicable models in 300, 400, 500 and 600 sizes with A, D and V voltage types. It specifies an ambient operating range of 15°C to 32°C, altitude up to 2000 m, maximum relative humidity of 60%, and voltage fluctuation limits. It also gives maintenance tasks such as condenser and filter cleaning, gasket maintenance, defrosting, battery maintenance and a maintenance schedule.
| Applicable sizes | 300, 400, 500, 600 |
|---|---|
| Voltage types | A, D, V |
| Ambient operating temperature | 15°C to 32°C (59°F to 90°F) |
| Maximum altitude | 2000 m |
| Maximum relative humidity | 60% |
| Voltage fluctuation (115 V/60 Hz and 230 V/50 Hz) | ±10% of nominal voltage |
| Voltage fluctuation (208-230 V/60 Hz) | -10% or +6% of nominal voltage |
| Operating environment classification | Pollution Degree 2, Overvoltage Category II |
| Current rating (300A) | 15.25 A at 115 V, 60 Hz |
| Current rating (300D) | 7.2 A at 208-230 V, 60 Hz |
| Current rating (300V) | 7.1 A at 230 V, 50 Hz |
| Current rating (400A) | 16.7 A at 115 V, 60 Hz |
| Current rating (400D) | 7.6 A at 208-230 V, 60 Hz |
| Current rating (400V) | 6.5 A at 230 V, 50 Hz |
| Current rating (500A) | 16.7 A at 115 V, 60 Hz |
| Current rating (500D) | 7.6 A at 208-230 V, 60 Hz |
| Current rating (500V) | 6.5 A at 230 V, 50 Hz |
| Current rating (600A) | 16.7 A at 115 V, 60 Hz |
| Current rating (600D) | 7.6 A at 208-230 V, 60 Hz |
| Current rating (600V) | 6.5 A at 230 V, 50 Hz |
Allow the freezer to operate at the desired temperature for a minimum of 12 hours before loading.
The refrigeration system is designed for an ambient environment within 15°C to 32°C (59°F to 90°F).
No, the manual states this unit is not a “rapid-freeze” device.
Keep the packing materials and immediately report the damage to the carrier.