This manual introduces the Keysight U1251B and U1252B handheld digital multimeters.
Keysight U1251B and U1252B Handheld Digital Multimeter Quick Start Guide covers initial setup and basic operation of the U1251B and U1252B multimeters. It lists included accessories, safety and EMC information, and battery charging steps for the rechargeable U1252B model.
The guide also summarizes front-panel functions, input terminals and overload protection, and procedures for voltage and current measurements, including AC/DC voltage, the filter for mixed signals, and AC current measurement.
| Overvoltage protection | CAT III 1000 V / CAT IV 600 V |
|---|---|
| Pollution degree | II |
| Voltage input overload protection | 1000 V R.M.S (COM) |
| Diode input overload protection | 1000 V R.M.S, <0.3 A short circuit current |
| Current input (µA/mA) overload protection | 440 mA/1000 V, 30 kA fast-acting fuse |
| Current input (A) overload protection | 11 A/1000 V, 30 kA fast-acting fuse |
| U1251B battery | 9 V alkaline |
| U1252B battery | Rechargeable 8.4 V |
| Charging supply | 24 V DC adapter |
The package includes test leads, test probes, alligator clips, a printed Quick Start Guide, a Certificate of Calibration, and a battery; the U1252B also includes a power cord and AC adapter.
Remove and disconnect the test leads, set the rotary switch to the charging position, connect the 24 V DC adapter, plug its terminals to V and COM, then press to start charging.
Enter Setup mode, scroll to FiLtE, enable the Filter, save the changes, then restart the meter.
Voltage and diode inputs are protected to 1000 V R.M.S.; the µA/mA input is protected to 440 mA/1000 V with a 30 kA fast-acting fuse; the A input is protected to 11 A/1000 V with a 30 kA fast-acting fuse.