Keysight U1251B and U1252B Handheld Digital Multimeter Quick Start Guide

This manual introduces the Keysight U1251B and U1252B handheld digital multimeters.


Keysight U1251B and U1252B Handheld Digital Multimeter Quick Start Guide - cover page
Brand
Keysight
Category
Other Electronics
Document type
Quick Start Guide
Language
English
Pages
177
File format
PDF
File size
6.8 MB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
5EDB47A41224D173F8D07B56BBC3668C

About this manual

Keysight U1251B and U1252B Handheld Digital Multimeter Quick Start Guide covers initial setup and basic operation of the U1251B and U1252B multimeters. It lists included accessories, safety and EMC information, and battery charging steps for the rechargeable U1252B model.

The guide also summarizes front-panel functions, input terminals and overload protection, and procedures for voltage and current measurements, including AC/DC voltage, the filter for mixed signals, and AC current measurement.

What's inside

  • Safety and EMC Information
  • Safety Notices
  • Charging the Battery
  • Functions and Features
  • Input Terminals and Overload Protection
  • Performing Voltage Measurements
  • Measuring AC voltage
  • Measuring DC voltage
  • Using the Filter
  • Performing Current Measurements

Specifications

Overvoltage protectionCAT III 1000 V / CAT IV 600 V
Pollution degreeII
Voltage input overload protection1000 V R.M.S (COM)
Diode input overload protection1000 V R.M.S, <0.3 A short circuit current
Current input (µA/mA) overload protection440 mA/1000 V, 30 kA fast-acting fuse
Current input (A) overload protection11 A/1000 V, 30 kA fast-acting fuse
U1251B battery9 V alkaline
U1252B batteryRechargeable 8.4 V
Charging supply24 V DC adapter

Frequently asked questions

What is included with the U1251B and U1252B multimeter?

The package includes test leads, test probes, alligator clips, a printed Quick Start Guide, a Certificate of Calibration, and a battery; the U1252B also includes a power cord and AC adapter.

How do I charge the battery?

Remove and disconnect the test leads, set the rotary switch to the charging position, connect the 24 V DC adapter, plug its terminals to V and COM, then press to start charging.

How do I enable the filter for DC voltage measurements?

Enter Setup mode, scroll to FiLtE, enable the Filter, save the changes, then restart the meter.

What are the input terminal overload protection limits?

Voltage and diode inputs are protected to 1000 V R.M.S.; the µA/mA input is protected to 440 mA/1000 V with a 30 kA fast-acting fuse; the A input is protected to 11 A/1000 V with a 30 kA fast-acting fuse.