Thermo Fisher Scientific Axiom Propel XPRES 384HT Workflow Site Preparation Guide

This document is a SITE PREPARATION GUIDE for use with Axiom™ 384HT Array Plates, Axiom™ Propel XPRES Reagent Kit, 2x384HT Multidrop™ Combi Reagent Dispenser


Thermo Fisher Scientific Axiom Propel XPRES 384HT Workflow Site Preparation Guide - cover page
Brand
Thermo Fisher Scientific
Category
Other
Document type
Installation Guide
Language
English
Pages
30
File format
PDF
File size
967 KB
Published
03 November, 2023
Updated
03 November, 2023
MD5 checksum
8F152C7307370B6D2D64795966F69D64

About this manual

The Axiom Propel XPRES 384HT Workflow Site Preparation Guide (Publication MAN0018488, Revision B.0) is a Thermo Fisher Scientific document for preparing a laboratory to run the Axiom Propel XPRES 384HT workflow. It lists arrays, reagents, software, equipment, labware, and accessories required, including the GeneTitan Multi-Channel Instrument, Multidrop Combi Reagent Dispenser, plate sealer, centrifuges, ovens, shakers, liquid handler, thermal cycler, and spectrophotometer.

The guide also covers reagent kit components and storage, a preamplification/amplification staging area, training kits, an example laboratory layout for 160K samples/year (8 plates/week) throughput, storage space requirements, and documentation and support. It states the products are for research use only and not for diagnostic procedures.

What's inside

  • Assay equipment and supplies required
  • Arrays, reagents, and software required
  • Axiom Propel XPRES Reagent Kit components
  • Equipment required
  • Preamplification/amplification staging area
  • GeneTitan Multi-Channel Instrument
  • Microplate dispenser
  • Plate sealer
  • Labware and accessories required
  • Laboratory layout example for 160K samples/year (8 plates/week) throughput

Specifications

Reagent kit capacitySufficient to process two 384HT array format plates
Required GeneTitan MC Instrument softwareGeneChip Command Console version 6.1.1 or later
Plate sealer air requirement50 L/min at 80–87 psi
Ionizing air blower discharge specification2 seconds at 12 in (30 cm), fan high

Frequently asked questions

What software version must the GeneTitan MC Instrument run?

It must run GeneChip Command Console software version 6.1.1 or later.

How many 384HT array plates can one Axiom Propel XPRES Reagent Kit process?

One kit is sufficient to process two 384HT array format plates.

What are the air requirements for the ALPS 3000 Automated Microplate Heat Sealer?

It requires 50 L/min at 80–87 psi.

Where should genomic DNA manipulations be performed?

The guide recommends a dedicated preamplification room or an area separate from the main laboratory, with dedicated pipettes and plasticware.