JENCO 6219 pH/ORP/Ion/Temperature Benchtop Tester User Manual

6219 is a multi-functional measurement instrument used in the laboratory, which can measure pH, temperature, oxidation-reduction potential and direct reading ion concentration.


JENCO 6219 pH/ORP/Ion/Temperature Benchtop Tester User Manual - cover page
Brand
JENCO
Category
Analyzer
Document type
User Manual
Language
Chinese
Pages
15
File format
PDF
File size
109 KB
Published
31 October, 2023
Updated
31 October, 2023
MD5 checksum
79E954C6938FC6B95FB3642C2D041BC1

About this manual

This is the Chinese user manual for the JENCO MODEL 6219, a microprocessor-based benchtop meter for laboratory use that measures pH, ORP (mV), ion concentration and temperature. It describes the two LCD displays, the eight keys, the connectors, automatic or manual temperature compensation, the low-battery (LOBAT) indication and the RS-232C link to a computer.

The manual gives the main technical specifications, then walks through battery replacement, calibration and measurement: preparing buffer solutions, two-point pH calibration with 6.86 and 4.00/9.18 buffers, pH measurement with ATC or manual temperature compensation, temperature measurement with the ATC/TEMP probe, ORP measurement, and the RS-232C settings and software notes. This manual is written in Chinese.

What's inside

  • Overview
  • Main technical specifications
  • Display description
  • Connectors
  • Keypad description
  • Operating steps
  • pH calibration
  • pH measurement
  • Temperature measurement
  • ORP measurement

Specifications

pH / mV measuring range-2.000 to 16.000 pH; -1999 to 1999 mV
Temperature range-5.0 °C to 125.0 °C
Ion concentration range0.0 to 119.8 CONC
Resolution0.001 pH; 0.1 mV; 0.1 °C
Input impedance≥1*10^12 Ω
Dimensions250 x 240 x 100 mm
Weight1.07 kg
Power supplySix AA batteries or UL-listed AC adapter (9 VDC, 100 mA)
Operating conditionsAmbient temperature 5 to 40 °C; relative humidity ≤85%
Calibration buffers4.00 pH, 6.86 pH, 9.18 pH
RS-232C baud rate2400 (fixed); 8 data bits, 1 stop bit

Frequently asked questions

Which standard buffer solutions does the 6219 use for calibration?

It uses 4.00 pH, 6.86 pH and 9.18 pH standard buffer solutions; the zero point is set in 6.86 pH and the slope in 4.00 or 9.18 pH.

How is the MODEL 6219 powered?

It runs on six AA batteries or from a UL-listed AC adapter rated 9 VDC, 100 mA.

Can the 6219 be connected to a computer?

Yes, it communicates through an RS-232C interface (DB-25 connector) so data can be observed and recorded on a computer.

What are the RS-232C settings of the instrument?

The transmission rate is fixed at 2400, data unit 8, stop unit 1, with the protocol controlled by the driver software.

Download PDF (109 KB)