6219 is a multi-functional measurement instrument used in the laboratory, which can measure pH, temperature, oxidation-reduction potential and direct reading ion concentration.
This is the Chinese user manual for the JENCO MODEL 6219, a microprocessor-based benchtop meter for laboratory use that measures pH, ORP (mV), ion concentration and temperature. It describes the two LCD displays, the eight keys, the connectors, automatic or manual temperature compensation, the low-battery (LOBAT) indication and the RS-232C link to a computer.
The manual gives the main technical specifications, then walks through battery replacement, calibration and measurement: preparing buffer solutions, two-point pH calibration with 6.86 and 4.00/9.18 buffers, pH measurement with ATC or manual temperature compensation, temperature measurement with the ATC/TEMP probe, ORP measurement, and the RS-232C settings and software notes. This manual is written in Chinese.
| pH / mV measuring range | -2.000 to 16.000 pH; -1999 to 1999 mV |
|---|---|
| Temperature range | -5.0 °C to 125.0 °C |
| Ion concentration range | 0.0 to 119.8 CONC |
| Resolution | 0.001 pH; 0.1 mV; 0.1 °C |
| Input impedance | ≥1*10^12 Ω |
| Dimensions | 250 x 240 x 100 mm |
| Weight | 1.07 kg |
| Power supply | Six AA batteries or UL-listed AC adapter (9 VDC, 100 mA) |
| Operating conditions | Ambient temperature 5 to 40 °C; relative humidity ≤85% |
| Calibration buffers | 4.00 pH, 6.86 pH, 9.18 pH |
| RS-232C baud rate | 2400 (fixed); 8 data bits, 1 stop bit |
It uses 4.00 pH, 6.86 pH and 9.18 pH standard buffer solutions; the zero point is set in 6.86 pH and the slope in 4.00 or 9.18 pH.
It runs on six AA batteries or from a UL-listed AC adapter rated 9 VDC, 100 mA.
Yes, it communicates through an RS-232C interface (DB-25 connector) so data can be observed and recorded on a computer.
The transmission rate is fixed at 2400, data unit 8, stop unit 1, with the protocol controlled by the driver software.