This manual is the operating manual for pH/ORP electrodes and references electrodes Memosens digital electrodes and analog electrodes, which contains basic safety guidelines, acceptance and product identification, installation, electrical connections, etc.
This operating manual describes Endress+Hauser pH/ORP electrodes and reference electrodes, including Memosens digital and analog versions. It lists intended uses for CPSx1D, CPSx1, CPS1x1D, CPSx2D, CPSx2, CPSx4, CPSx5 and CPSx3 electrodes.
The document covers document information, basic safety instructions, storage and transport, certificates and approvals, installation direction, electrical connection with CYK10/CYK20, CPK9 and CPK1 cables, commissioning and calibration, maintenance, sensor cleaning and regeneration of slow-response pH electrodes. It also references safety guidelines XA00376C and XA00028C for use in hazardous areas. This manual is written in Chinese.
| Storage temperature | 10...30 °C (50...90 °F) |
|---|---|
| Minimum horizontal installation angle | 15° |
| Lower temperature limit before electrode breakage | -5 °C (23 °F) |
| Storage solution when not using the protective cap | KCl solution (3 mol/l) or buffer |
| Soaking time after long dry storage | At least 24 hours in the medium |
| pH electrode calibration | Two-point calibration |
| ORP electrode calibration | Single-point calibration |
They must be stored in a dry environment at 10...30 °C (50...90 °F).
Yes. pH electrodes require two-point calibration.
Place it in the medium for at least 24 hours before use, otherwise severe drift can occur.
They are cleaned with a hydrofluoric acid mixture containing 10% nitric acid and 50 g/l ammonium fluoride.