9 color human white blood cell immunophenotyping, 13 color lymphocyte subset immunophenotyping, BrdU cell proliferation, CBA, etc.
This application manual for the Beckman Coulter CytoFLEX 524C flow cytometer describes experimental methods for multicolor immunophenotyping and cell function analysis using flow cytometry, intended for immunology research and related cell detection.
It covers protocols such as 9-color leukocyte immunophenotyping, 13-color lymphocyte subset analysis, BrdU and CFSE proliferation, cell cycle, apoptosis, microparticle and sperm DNA integrity testing, with sample preparation, instrument operation, gating strategy and result analysis. This manual is written in Chinese.
| Laser wavelengths | 405 nm, 488 nm, 638 nm |
|---|---|
| Fluorescence detection channels | 13 |
It uses 405 nm, 488 nm and 638 nm lasers.
At least 20000 events are acquired within the Lymph gate.
An αt index of 0.08 indicates good sperm integrity.