SONBEST SM3568B RS485 Irradiance Sensor User Manual

SM3568B uses industrial standard RS485 bus MODBUS-RTU protocol interface, which is convenient for connecting to PLC, DCS and other instruments or systems to monitor radiation and other state quantities. The internal use of high-precision sensor core and related devices ensure that the product has high reliability and excellent long-term stability


SONBEST SM3568B RS485 Irradiance Sensor User Manual - cover page
Brand
SONBest
Category
sensor
Document type
User Manual
Language
Chinese
Pages
11
File format
PDF
File size
641 KB
Published
30 October, 2023
Updated
30 October, 2023
MD5 checksum
21723E8995BBAE6C995DC2E591E38FC8

About this manual

The SM3568B is an RS485 irradiance sensor from SONBEST (Shanghai Sonbest Industrial Co., Ltd.). It uses MODBUS-RTU over an RS485 interface and is intended for connection to PLCs, DCS systems, and other instruments to monitor irradiance.

The manual lists a 0~2000 W/m2 irradiance range, 380~730 nm wavelength range, DC9~24V power, and RS485 communication at a default 9600 8 n 1. It also provides wiring guidance, a shipping list, MODBUS-RTU read commands, and instructions for changing the device address, baud rate, and correction value. This manual is written in Chinese.

What's inside

  • Technical Parameters
  • Product Wiring
  • Typical Application
  • Shipping List
  • RS485 Communication Protocol
  • Common Data Address Table
  • Read and Modify Device Address
  • Read and Modify Baud Rate
  • Read and Modify Correction Value
  • Contact Us

Specifications

Irradiance measurement range0~2000 W/M2
Maximum allowable irradiance error±7%
Irradiance repeatability±5%
Irradiance sensor elementImported digital
Wavelength measurement range380nm~730nm
Communication interfaceRS485
Default baud rate9600 8 n 1
Power supplyDC9~24V 1A
Operating temperature-30~85℃
Operating humidity5%RH~90%RH

Frequently asked questions

What is the default device address and baud rate?

The factory default device address is 1, and the default baud rate is 9600, 8, n, 1 for modules and non-recording instruments.

What protocol does the SM3568B use for communication?

It uses the RS485 MODBUS-RTU standard protocol, with all operation and reply commands in hexadecimal data.

What is the irradiance measurement range of the SM3568B?

The irradiance measurement range is 0 to 2000 W/m2.

How can I query the device address if I do not know it?

Use address 250 (FA) as the universal address with the command FA 03 00 66 00 01 71 9E to read the real device address.