This is an instruction manual for a SUN TR-SUN23-III autoclave, including product application scope, technical parameters, diagram specification, installation instructions, display and keypad description, emergency situation, maintenance and service, transportation and storage, attention and accessories
The SUN TR-SUN23-III is a microprocessor-controlled autoclave designed for frequent sterilization in dental, ophthalmology, and laboratory settings. It features an LCD display, automatic malfunction detection, and safety protections against excessive temperature and pressure.
The manual includes technical parameters, installation requirements, display and keypad operation, sterilization programs, water filling and draining procedures, error code troubleshooting, maintenance tasks such as fuse and sealing ring replacement, and transportation guidelines.
| Inner chamber volume | 18L (245mm x 355mm) / 22L (245mm x 465mm) |
|---|---|
| Rated voltage | 220V±22V, 50Hz±1Hz / 110V±11V, 60Hz±1Hz |
| Rated power | 1800VA / 2000VA |
| Sterilizing pressure/temperature | 90-130KPa/121ºC and 200-230KPa/134ºC |
| Fuse rating | T15A / T30A |
| Water tank capacity | 4.5L |
| Water consumption per cycle | 0.3-0.5L |
| Operating temperature range | 0-40ºC |
The autoclave offers seven programs: 134°C unwrapped, 121°C unwrapped, 134°C wrapped, 121°C wrapped, 134°C cotton, 121°C cotton, and 121°C plastic, each with specific pressure, vacuum cycles, sterilizing time, and drying time.
After turning on the power switch, fill approximately 3000ml of distilled water from the top until near the red mark; do not overfill.
E6 indicates a problem with the door switch or the door being opened during operation; check that the door handle is rotated to the maximum position.
The sealing ring should be cleaned periodically with a soft cloth and distilled water; if air leakage persists, remove the ring for deeper cleaning or replacement.