This guide describes how to use the VeTMAX™ C. burnetii & Chlamydophila spp. Kit for nucleic acid purification
This document outlines nucleic acid purification protocols optimized for use with the VetMAX C. burnetii & Chlamydophila spp. Kit (Cat. No. TFQQCHP). The protocols cover sample selection, storage, and preparation for various matrices including organs, swabs, fluids, and milk.
The guide details procedures for both automated nucleic acid purification using instruments like the KingFisher Flex, MagMAX Express-96, and KingFisher Duo Prime, as well as manual purification using silica-based spin columns such as the QIAamp DNA Mini Kit and NucleoSpin Tissue kit.
| Species | Bovine, Small ruminants (sheep, goats) |
|---|---|
| Sample matrices | Organ, Swab, Fetal fluid, Vaginal mucus, Milk |
| Organ sample quantity | 20-30 mg of tissue |
| Swab sample quantity | Individual |
| Fluid sample quantity | 200 µL of fluid |
| Milk sample quantity | 200 µL of milk |
| Organ storage (short term) | 2°C to 8°C if analysis within 24 hours |
| Organ storage (long term) | Below −16°C for up to 1 year |
Automated purification is performed using the KingFisher Flex, MagMAX Express-96, KingFisher mL, or KingFisher Duo Prime instruments.
Store samples below −16°C if the analysis is to be performed more than 24 hours after sampling.