This document is about the user manual for W-K-F,W-M-F temperature sensors. It mainly introduces the product description, usage temperature range, technical data, installation instructions, type label and other information
The document provides a user manual and installation guide for Lapp EPIC SENSORS temperature sensors, specifically models W-K-F (plastic enclosure) and W-M-F (metal enclosure). These devices are designed for measuring room or ambient temperatures ranging from -40°C to +80°C using a Pt100 resistance sensor. The sensors feature an ingress protection rating of IP 65 and are intended for professional installation on walls or various structures. Metal versions are also available with Ex i (intrinsically safe) and Ex d (flameproof) certifications for hazardous areas.
The guide details the product description, technical data, materials used, and dimensional drawings. It includes step-by-step installation instructions, such as drilling holes, securing the enclosure, and connecting cables. Specific sections cover opening the enclosure for standard and Ex versions, wiring Pt100 connections according to EN 60751 standards, and calculating measuring currents to prevent self-heating errors. Additionally, the manual provides extensive data on Ex i specifications, including maximum interface values, allowed ambient temperatures, and thermal resistance calculations for self-heating in hazardous environments.
| Measuring range (Pt100) | -40…+80 °C |
|---|---|
| Ingress protection rating | IP 65 |
| Enclosure material (Plastic) | Polycarbonate (PC-V0) |
| Enclosure material (Metal) | Aluminum |
| Sensor element material | AISI 316L |
| Max measuring current (Pt100) | 1 mA |
| Max measuring current (Pt500) | 0.5 mA |
| Max measuring current (Pt1000) | 0.3 mA |
The standard measuring range for the Pt100 sensor is -40°C to +80°C.
Yes, metal enclosure versions (W-M-F) are available with Ex i and Ex d certifications for hazardous areas.
The normally recommended maximum measuring current for a Pt100 resistor is 1 mA.
Open the enclosure cover by opening four (4) cover screws, one on each corner.