Thermo Fisher Scientific Ion S5 / Ion GeneStudio S5 Sequencing Run Report Assessment User Bulletin

This document describes an approach to evaluate an Ion S5 ™ or Ion GeneStudio ™ S5 sequencing run report. Although each sequencing run should be evaluated as a whole, there are individual metrics that can help you assess a sequencing run. This document provides an overall view of expected run performance. For cases that involve further help, contact Thermo Fisher Scientific Technical Support at www.thermofisher.com/support, or your contract support personnel.


Thermo Fisher Scientific Ion S5 / Ion GeneStudio S5 Sequencing Run Report Assessment User Bulletin - cover page
Brand
Thermo Fisher Scientific
Category
Other
Document type
Other
Language
English
Pages
13
File format
PDF
File size
1.3 MB
Published
26 October, 2023
Updated
26 October, 2023
MD5 checksum
102CC219646D9906D546B35897C94E87

About this manual

This user bulletin from Thermo Fisher Scientific describes an approach to evaluate an Ion S5 or Ion GeneStudio S5 sequencing run report. It explains that each sequencing run should be evaluated as a whole, while individual metrics can help assess a sequencing run, and it directs users to Technical Support for further help.

The document covers pre-alignment and post-alignment metrics, throughput considerations for Ion 510–550 chips, read filtering and trimming, ISP loading density, and factors that can affect loading density, including template preparation and bubbles. An example run report from a 200-base read run on an Ion 540 Chip is included, and Torrent Suite Software 5.10 is used to view the report.

What's inside

  • Run report metrics before alignment
  • Run report metrics after alignment
  • Throughput considerations
  • Read filtering and trimming
  • ISP loading density
  • Factors that can affect loading density
  • Variable images
  • Bubbles

Specifications

Ion 510 Chip addressable wells and reads~6 million wells; 2–3 million reads
Ion 520 Chip addressable wells and reads~12 million wells; 4–6 million reads
Ion 530 Chip addressable wells and reads~37 million wells; 15–20 million reads
Ion 540 Chip addressable wells and reads~150 million wells; 60–80 million reads
Ion 550 Chip addressable wells and reads~260 million wells; 100–130 million reads

Frequently asked questions

What metrics are reviewed before alignment?

The document describes pre-alignment metrics for unaligned reads, including total bases, key signal, ISP loading, total reads, usable reads, and read length.

What does the usable reads percentage measure?

It is the percentage of library ISPs that pass the polyclonal, low quality, and primer-dimer filters.

What is ISP loading density?

It is the percentage of chip wells that contain an Ion Sphere Particle, considering only addressable wells.

What is the final library percentage?

It is the percentage of reads that pass all filters and are recorded in the output BAM file.