This manual describes the features, specifications, applications, accessories, and shipping of the 6108 and 610BR Electrometers.
This manual provides detailed instructions for operating the Keithley 6108 and 610BR electrometers, including voltage, current, resistance, and charge measurements. It covers calibration procedures, front and rear panel controls, and accessory usage. The 610BR is a rack-mounted version of the 6108 with identical circuitry and specifications, except for front-panel controls. Both models are factory-calibrated and shipped with essential components and documentation.
| Voltage Range | 0.001 volt full scale to 100 volts |
|---|---|
| Current Range | 10-14 ampere full scale to 0.3 ampere |
| Resistance Range | 100 ohms full scale to 1014 ohms |
| Charge Range | 10-12 coulomb full scale to 10-5 coulomb |
| Voltage Accuracy | ±1% of full scale |
| Zero Drift | 200 microvolts per hour max. averaged over any 24-hour period after warm-up |
| Input Impedance | Greater than 10^14 ohms shunted by 22 picofarads |
The 610BR is a rack-mounted version of the 6108 with identical circuits and specifications, except that its COARSE ZERO Control is on the front panel, not the rear panel.
Input impedance is greater than 10^14 ohms shunted by 22 picofarads on the VOLTS position, selectable in decade steps from 10 to 10^14 ohms.