About this manual

The PDP90A is a two-dimensional, tetra-lateral position sensing detector designed for precise alignment and position measurements. It works with Thorlabs’ TPA101 T-Cube Auto-Aligner and supports spot sizes up to 9 mm in diameter. Setup involves mounting the sensor, connecting it to the controller, and adjusting beam intensity to ensure optimal signal-to-noise ratio. The sensor provides X and Y-axis positional data via four electrodes and includes circuitry to decode position signals.

Technical specifications include a wavelength range of 320–1100 nm, peak responsivity of 0.6 A/W at 960 nm, and a maximum photocurrent of 40 µA. The sensor operates within a temperature range of 10–40°C and is designed for applications requiring precise alignment, such as microscope focusing and fiber launch systems.

What's inside

  • Warning Symbol Definitions
  • Description
  • Setup Instructions
  • Details of Operation
  • Calculating the Position
  • Position Resolution
  • Specifications
  • Regulatory
  • Thorlabs Worldwide Contacts

Specifications

Wavelength Range320 – 1100 nm
Peak Responsivity0.6 A/W at 960 nm
Photocurrent (Max)40 µA
Output Voltage Range0 to 4 V
Spot SizeØ0.2 to Ø7 mm (Recommended), Ø9 mm (Max)
Operating Temperature10 to 40 °C
Sensor Size9 mm x 9 mm
Mechanical ApertureØ0.50” (Ø12.7 mm)
Weight0.25 lbs. (114 g)

Frequently asked questions

What is the maximum spot size supported by the PDP90A?

The maximum spot size is Ø9 mm, though a spot between Ø0.2 mm and Ø7 mm is recommended for best results.

What is the position resolution of the PDP90A?

The position resolution is 0.75 µm when the SUM output voltage is set to 4 V.

What is the maximum photocurrent of the PDP90A?

The maximum photocurrent is 40 µA.

What is the operating temperature range for the PDP90A?

The operating temperature range is 10 to 40 °C.